HI 93751
Sulfate Photometer with 470 nm LED
Sulfate is widely present in natural waters in different concentrations.
Sulfate concentration is to be kept within a strict range for drinking water, especially since the value can be high near mine drainage points
Sulfate is also rigorously tested in the production of beverages such as beer, due to its significant effect upon odor and taste.
Sulfate is added to certain types of boilers to help precipitate calcium and magnesium and to inhibit encrustation.
On the other hand, too much sulfate can be corrosive in high pressure boilers, electric turbines and their heat exchangers.
In fact, in these applications it is important to keep the level below a specific limit.
Similar checks of sulfate presence are carried out in water used for different production cycles, including those of semiconductors.
HI 93751 measures the sulfate ion concentration using a turbidimetric method. The turbidity of the sample, after the addition of reagents, is proportional to the strength of sulfate present.
Order Information:
HI 93751 is supplied with 2 cuvets, battery and instructions.
Specifications
Resolution | 1 mg/L | |
Accuracy | ±1 mg/L ±5% of reading | |
Light Source | LED 470 nm | |
Light Life | Life of the instrument | |
Light Detection | Silicon Photocell | |
Battery Type / Life | 1 x 9V / approx. 40 hours of continuous use; auto-off after 10 minutes of non use | |
Environment | 0 to 50°C (32 to 122°F); RH max 95% non-condensing | |
Dimensions | 180 x 83 x 46 mm (7.1 x 3.3 x 1.8″) | |
Weight | 290 g (10 oz.) | |
Method | Adaptation of the turbidimetric method. The reaction between sulfate and the reagent causes turbidity in the sample |
Solutions | |
HI 93703-50 | Solution for Cleaning Cuvets |
Accessories | |
HI 731321 | Glass Cuvet for HI 937xx & HI 832xx Series Meters |
HI 731325 | Cuvet Cap Turbidity Meters & HI 937xx Series Meters |
HI 710009 | Shockproof Boot, Blue |
HI 710010 | Shockproof Boot, Orange |
HI 731318 | Tissue for Wiping Cuvets |
Reagents & Standards | |
HI 93751-01 | Sulfate, Turbidimetric Method (100 Tests) |
HI 93751-03 | Sulfate, Turbidimetric Method (300 Tests) |