Product Features
| :: | High Resolution and Accuracy (Up to ±0.002 pH Accuracy) for precise measurement |
| :: | Backlit Two Line, Dot Matrix LCD Display provides clear instructions for setup, measurement and calibration |
| :: | Direct Ion Concentration measurement with use of many Ion Selective Electrodes (Only Available in CyberScan pH 2100) |
| :: | Multi-Point Push-Button Calibration with Auto-Buffer Recognition (Up to 5 points) for optimum full-range accuracy |
| :: | User-Customization for enhanced flexibility like pH resolution, etc. and viewing of electrode diagnosis – pH slope/offset/mV base value, and previous calibration points |
| :: | Calibration Due Alarm prompts at user-defined interval to ensure regularly scheduled recalibration for ISO and GLP compliance |
| :: | Automatic Downloading of data into memory or computer via a cable at desired time intervals or whenever reading stabilizes |
| :: | Expanded Memory stores and recalls up to 100 data sets either manually or on programmable real-time intervals |
| :: | Built-In Real-time Clock stamps stored data and calibration data with time & date – meets standards for GLP purposes |
| :: | High/Low Set Point Alarm alerts once reading outside preset limits |
| :: | RS232 Communication interface to download data to a printer or PC with complimentary Windows®-based DAS software for report documentation |
| :: | Analogue Recorder for external chart recording devices |
| :: | Automatic Temperature Compensation (ATC) for best accuracy under fluctuating temperatures |
| :: | Stability Criteria options determine stability of reading from instantaneous to moderate speed ideal for difficult sample measurement |
| :: | Hold Function momentarily freezes reading for easy viewing and recording |
| :: | Self-Diagnostic with message codes for easy troubleshooting |
| :: | Slide-Out Instruction Card for quick reference |
| :: | Detachable Electrode Stand organizes multiple electrodes neatly and holds them firmly in place |
Applications
Laboratory/Industrial: Use for water quality testing, sewage treatment, soil/fertilizer testing, food processing (for HACCP compliance), geological testing. Great for R&D and QC applications in the food and pharmaceutical industries, paper/pulp processing, chemical industries and related fields. Suitable for most ASTM and EPA test procedures.
Educational: Perfect for research labs, suitable for all types of chemistry labs and ecology studies.
ORP: Use ORP electrode to monitor plating operations, ozonation system efficiency, and other redox measurements.
Specifications
| pH Range | -1.000 to 14.000 pH |
| Resolution & Accuracy | 0.01/0.001 pH & ±0.01/0.002 pH |
| Ion Concentration Range | – |
| Resolution & Accuracy | – |
| mV Range | -1850 to +1850 mV |
| Resolution & Accuracy | 0.1 mV & ±0.2 mV |
| Temperature Range | -5.0 to 105.0 °C (23.0 to 221.0 °F) |
| Resolution & Accuracy | 0.1 °C (0.1 °F) & ±0.5 °C (0.9 °F) |
| No. of Calibration Points | 1 to 5 |
| pH Buffer Options | USA: 1.68, 4.01. 7.00, 10.01. 12.45 NIST: 1.68, 4.01, 6.86, 9.18, 12.45 User 1: 1.68, 4.01, 6.86, 10.01, 12.45 User 2: 1.68, 4.01, 6.86, 10.01, 12.45 Custom: min. of 2 and max. of 5 points |
| No. of Ion Calibration Points | – |
| Memory | 100 Data Sets with Date/Time stamping |
| Temperature Compensation | Automatic/Manual (0 to 100 °C) |
| pH Slope & Offset Display | Yes |
| Print Output | Bi-directional RS232C, Recorder (±1850 mV) |
| Display | Alpha-numeric Dot-matrix LCD with backlit |
| Special Functions | Auto-buffer recognition, Date/time indexed (GLP compliance); Calibration due prompt, Alarm; Stability criteria Stability Indicator & Self-diagnostic |
| Power | 9VDC, using AC adapters 110VAC or 220VAC, 50/60 Hz |